Performance Verification
1-17
4.
Set the decade resistance box to each of the values shown in
Table 1-8
, and verify that
the temperature readings are within the required limits.
Verifying frequency
Follow the steps below to verify the Model 2700 frequency function:
1.
Connect the function generator to the Model 2700 INPUT jacks. (See
Figure 1-7
.) Be
sure the INPUTS switch is in the FRONT position.
2.
Set the function generator to output a 1kHz, 1V RMS sine wave.
3.
Select the Model 2700 frequency function by pressing the FREQ key.
4.
Verify that the Model 2700 frequency reading is between 999.9Hz and 1.0001kHz.
Figure 1-7
Connections for Model 2700 frequency verification
Table 1-8
Four-wire RTD temperature verification reading limits
Applied resistance*
Reading limits (1 year, 18°C to 28°C)
22.80
Ω
100.00
Ω
313.59
Ω
-190.06 to -189.94°C
-0.06 to +0.06°C
599.94 to 600.06°C
*Based on
α
= 0.00385. See text.
Function Generator
Model 2700
Model 2700
Multimeter / Data Acquisition System
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE
RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
GPIB
BNC-to-Dual
Banana Plug
Adapter
50
Ω
Coax
Cable
Function
Output
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...