Routine Maintenance
Replacing the front terminal AMPS fuse
The front terminal AMPS fuse protects the Model 2700 current input from an over-current
condition. Follow the steps below to replace the AMPS fuse.
WARNING
Make sure the instrument is disconnected from the power line and other
equipment before replacing the AMPS fuse.
1.
Turn off the power, and disconnect the power line and test leads.
2.
From the front panel, gently push in the AMPS jack with your thumb, and rotate the
fuse carrier one-quarter turn counterclockwise. (See
Figure 3-2
.) Release pressure on
the jack, and its internal spring will push the fuse carrier out of the socket.
3.
Remove the fuse, and replace it with the same type: 3A, 250V, fast-blow, Keithley part
number FU-99-1.
CAUTION
Do not use a fuse with a higher current rating than specified or instrument
damage may occur. If the instrument repeatedly blows fuses, locate and
correct the cause of the trouble before replacing the fuse.
4.
Install the new fuse by reversing the above procedure.
Model 2700
Multimeter / Data Acquisition System
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
EXIT
ENTER
DIGITS RATE
REL
FILTER
TRIG
EX TRIG
STORE RECALL
OPEN
DCV
DCI
MATH
O U T P U T
RATIO
ACV
ACI
Ω
2
Ω
4
FREQ
TEMP
CH AVG
CONT
PERIOD SENSOR
LIMITS
ON/OFF
DELAY
HOLD
SAVE
SETUP
CONFIG
HALT
TYPE
LSYNC
TEST
MONITOR
STEP
SCAN
OCOMP
CH-OFF
CARD
CLOSE
Integra Series
RS-232
GPIB
Push in AMPS
and Rotate
Counterclockwise
Model 2700
Figure 3-2
Front terminal
AMPS fuse
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...