Calibration Reference
B-15
Manufacturing calibration commands
Three calibration steps are only performed at the factory or when the unit has been repaired:
:CALibration:PROTected:AC:STEP14
1V AC at 3Hz
:CALibration:PROTected:AC:STEP15
1V AC at 1kHz
:CALibration:PROTected:DC:STEP0
Rear scanner terminal short circuit
:AC:STEP<14|15>
(CALibration:PROTected:AC:STEP<14|15>)
Purpose
To program individual AC manufacturing calibration steps.
Format
:cal:prot:ac:step14 <Cal_voltage>
:cal:prot:ac:step15 <Cal_frequency>
Parameter
<Cal_voltage> = 1 [1V nominal]
<Cal_frequency = 1E3 [1kHz nominal]
Description
The
:AC:STEP14
and
:AC:STEP:15
commands program the two man-
ufacturing AC calibration steps. The appropriate signal must be connected
to the instrument when programming each step, as summarized by the
parameters listed above.
Example
:CAL:PROT:AC:STEP14 1
Program AC step 14.
:CAL:PROT:AC:STEP15 1E3
Program AC step 15.
:DC:STEP0
(:CALibration:PROTected:DC:STEP0)
Purpose
To perform rear scanner terminal short-circuit calibration.
Format
:cal:prot:dc:step0
Parameter
None
Description
:STEP0
performs the rear scanner terminal short-circuit calibration step in
the manufacturing calibration procedure. Install an extender card with low-
thermal shorts on the output terminals installed in Slot 1, and select the rear
scanner inputs before sending this command.
Example
:CAL:PROT:DC:STEP0
Perform rear short-circuit calibration.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...