4-16
Troubleshooting
Table 4-12
through
Table 4-16
can be used to trace the analog signal through the A/D multi-
plexer (U163) to the final amplifier stage. These tables show the MUX lines (S3, S4, S6, S7)
that are selected for measurement during the SIGNAL phase of the multiplexing cycle. Also
included are switching states of analog switches (U129) that set up the gain for the final ampli-
fier stage (U166).
Table 4-12
DCV signal multiplexing and gain
Range
Signal
(U163)
U129
pin 1
U129
pin 8
U129
pin 9
Gain
(U166)
100mV
1V
10V
100V
1000V
S4
S4
S4
S4
S4
OFF
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
ON
OFF
OFF
OFF
OFF
×
100
×
10
×
1
×
10
×
1
Table 4-13
ACV and ACA signal multiplexing and gain
Range
Signal
(U163)
U129
pin 1
U129
pin 8
U129
pin 9
Gain
(U166)
All
S3
ON
OFF
OFF
×
1
Table 4-14
DCA signal multiplexing and gain
Range
Signal
(U163)
U129
pin 1
U129
pin 8
U129
pin 9
Gain
(U166)
20mA
100mA
1A
3A
S6
S6
S6
S6
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
ON
ON
ON
OFF
×
100
×
100
×
100
×
10
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...