2-16
Calibration
DC current calibration
After the 1M
Ω
resistance point has been calibrated, follow these steps for DC current
calibration:
1.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2700 as
shown in
Figure 2-3
.
2.
Perform the calibration steps listed in
Table 2-10
. For each step:
•
Set the calibrator to the indicated current, and make sure the unit is in operate.
(Use the recommended current if possible.)
•
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
•
Wait until the Model 2700 completes each step before continuing.
NOTE
If you are performing DC-only calibration, proceed to “Programming calibration
dates.”
AC voltage calibration
Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2700 INPUT HI and LO terminals as shown in
Figure 2-4
.
2.
Perform the calibration steps summarized in
Table 2-11
. For each step:
•
Set the calibrator to the indicated voltage and frequency, and make sure the unit is
in operate. (You must use the stated voltage and frequency.)
•
Send the indicated programming command.
•
Wait until the Model 2700 completes each step before continuing.
Table 2-10
DC current calibration programming steps
Calibration
step
Calibrator
current
Calibration command*
Parameter range
10mA
100mA
1A
10.00000mA
100.00000mA
1.000000A
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1
*Change parameter if using different current.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...