Calibration
2-9
3.
Perform the steps listed in
Table 2-3
to complete DC volts calibration. For each calibra-
tion step:
•
Set the calibrator to the indicated value, and make sure it is in OPERATE.
•
Press the ENTER key to calibrate that step.
•
Wait until the Model 2700 finishes each step. (The unit will display the
CALIBRATING message while calibrating.)
NOTE
If your calibrator cannot output the values recommended in
Table 2-3
, use the left and
right arrow keys, and the up and down range keys to set the Model 2700 display value
to match the calibrator output voltage.
Resistance calibration
Completing the 100V DC calibration step ends the DC voltage calibration procedure. The
Model 2700 will then prompt you to connect 1k
Ω
. Follow these steps for resistance calibration:
1.
Set the calibrator output for resistance, and turn on external sense.
NOTE
Use external sense (4-wire
Ω
) when calibrating all resistance ranges. Be sure that
the calibrator external sense mode is turned on.
2.
Perform the calibration steps summarized in
Table 2-4
. For each step:
•
Set the calibrator to the indicated value, and place the unit in operate. (If the cali-
brator cannot output the exact resistance value, use the Model 2700 left and right
arrow keys and the range keys to adjust the Model 2700 display to agree with the
actual calibrator resistance.)
•
Press the ENTER key to calibrate each point.
•
Wait for the Model 2700 to complete each step before continuing.
Table 2-3
DC volts calibration summary
Calibration step
Calibrator voltage
Allowable range
+10V
-10V
100V
+10.00000V
-10.00000V
+100.0000V
+9V to +11V
-9V to -11V
+90V to +110V
Table 2-4
Ohms calibration summary
Calibration step
Calibrator resistance*
Allowable range
1k
Ω
10k
Ω
100k
Ω
1M
Ω
1k
Ω
10k
Ω
100k
Ω
1M
Ω
0.9k
Ω
to 1.1k
Ω
9k
Ω
to 11k
Ω
90k
Ω
to 110k
Ω
0.9M
Ω
to 1.1M
Ω
*Nominal resistance. Adjust Model 2700 calibration parameter to agree with actual value.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...