Calibration Reference
B-3
Command
Description
:CALibration
:PROTected
:DC
:STEP10 <NRf>
10mA DC step.
:STEP11 <NRf>
100mA DC step.
:STEP12 <NRf>
1A DC step.
:AC
AC cal steps.
:STEP1
10mV AC at 1kHz step.
:STEP2
100mV AC at 1kHz step.
:STEP3
100mV AC at 50kHz step.
:STEP4
1V AC at 1kHz step.
:STEP5
1V AC at 50kHz step.
:STEP6
10V AC at 1kHz step.
:STEP7
10V AC at 50kHz step.
:STEP8
100V AC at 1kHz step.
:STEP9
100V AC at 50kHz step.
:STEP10
700V AC at 1kHz step.
:STEP11
100mA AC at 1kHz step.
:STEP12
1A AC at 1kHz step.
:STEP13
2A AC at 1kHz step.
:STEP14
1V AC at 3Hz step
1
.
:STEP15
1V AC at 1kHz step
1
.
NOTE: Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.”
1
DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in manufacturing calibration
mode.
Table B-1 (Continued)
Remote calibration command summary
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...