![Keithley 2700 Service Manual Download Page 146](http://html1.mh-extra.com/html/keithley/2700/2700_service-manual_3944616146.webp)
B-10
Calibration Reference
:STEP5
(:CALibration:PROTected:DC:STEP5)
Purpose
To program the 100V DC comprehensive calibration step.
Format
:cal:prot:dc:step5 <Cal_voltage>
Parameter
<Cal_voltage> = 90 to 110 [V]
Description
:STEP5
programs the 100V DC comprehensive calibration step. The
allowable range of the calibration voltage parameter is from 90 to 110, but
100 is recommended for best results.
Example
:CAL:PROT:DC:STEP5 100
Program 100V step.
:STEP6
(:CALibration:PROTected:DC:STEP6)
Purpose
To program the 1k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step6 <Cal_resistance>
Parameter
<Cal_resistance> = 900 to 1.1E3 [
Ω
]
Description
:STEP6
programs the 1k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from 900
to 1.1E3 but 1E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP6 1E3
Program 1k
Ω
step.
:STEP7
(:CALibration:PROTected:DC:STEP7)
Purpose
To program the 10k
Ω
4-wire comprehensive calibration step.
Format
:cal:prot:dc:step7 <Cal_resistance>
Parameter
<Cal_resistance> = 9E3 to 11E3 [
Ω
]
Description
:STEP7
programs the 10k
Ω
4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
9E3 to 11E3, but 10E3 is recommended for best results.
Example
:CAL:PROT:DC:STEP7 10E3
Program 10k
Ω
step.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...