Performance Verification
1-19
2.
Install the Model 7700 in Slot 1 of the Model 2700, then turn on the power, and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Select the DC volts function by pressing the
DCV
key, and set the Model 2700 to the
100mV range. Close Channel 1 by pressing the CLOSE key and then keying in 101.
4.
Set the calibrator output to 0.00000mV DC, and allow the reading to settle.
5.
Enable the Model 2700 REL mode. Leave REL enabled for the remainder of the DC
volts verification tests.
6.
Source positive and negative and full-scale voltages for each of the ranges listed in
Table 1-9
. For each voltage setting, be sure that the reading is within stated limits.
7.
Press the OPEN key to open Channel 1.
Table 1-9
Plug-in module DCV reading limits
Range
Applied DC voltage*
Reading limits (1 year, 18° to 28°C)
100mV
1V
10V
100V
1000V
100.0000mV
1.000000V
10.00000V
100.0000V
300.000V
99.9935 to 100.0065mV
0.999963 to 1.000037V
9.99965 to 10.00035V
99.9946 to 100.0054V
299.976 to 300.024V
*Source positive and negative values for each range.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...