Calibration Reference
B-5
:INIT
(:CALibration:PROTected:INITiate)
Purpose
To initiate comprehensive and factory calibration procedures.
Format
:cal:prot:init
Parameter
None
Description
The
:INIT
command enables Model 2700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the
:CODE
command, but before sending any other calibration
command.
NOTE
The
:INIT
command should be sent only once before performing either DC, AC, or
factory calibration. Do not send
:INIT
before each calibration step.
Example
:CAL:PROT:INIT
Initiate calibration.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
To lock out comprehensive or manufacturing calibration.
Format
:cal:prot:lock
Parameter
None
Description
The
:LOCK
command allows you to lock out both comprehensive and man-
ufacturing calibration after completing those procedures. Thus,
:LOCK
per-
forms the opposite of enabling calibration with the
:CODE
command.
NOTE
To unlock comprehensive calibration, send the
:CODE
command. To unlock manu-
facturing calibration, hold in the OPEN key while turning on the power.
Example
:CAL:PROT:LOCK
Lock out calibration.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...