![Keithley 2700 Service Manual Download Page 131](http://html1.mh-extra.com/html/keithley/2700/2700_service-manual_3944616131.webp)
Specifications
A-11
GENERAL
RELAY SWITCH CONFIGURATION: 40 independent channels of 1-
pole switching. Isolated from internal DMM.
CONTACT CONFIGURATION: 1 pole Form A.
RELAY TYPE: Latching electromechanical.
CONNECTOR TYPE: Two 50-pin female D-sub connectors.
INPUTS
MAXIMUM SIGNAL LEVEL: 300VDC or rms, 2A switched, 60W (DC,
resistive), 125VA (AC, resistive).
CONTACT LIFE: Cold Switching: 10
8
closures.
At Maximum Signal Levels: 10
5
closures.
CHANNEL RESISTANCE (per conductor): <1
Ω
.
CONTACT POTENTIAL:
≤
4µV per contact.
OFFSET CURRENT: <100pA.
ACTUATION TIME: 3ms.
ISOLATION: Channel to Channel: >10
9
Ω
, <50pF.
Common Mode: >10
9
Ω
, <100pF.
CROSSTALK (1MHz, 50
Ω
load): <–35dB.
INSERTION LOSS (50
Ω
source, 50
Ω
load): <0.3dB below 1MHz, <3dB
below 10MHz.
COMMON MODE VOLTAGE: 300V between any terminal and chassis.
ENVIRONMENTAL
OPERATING ENVIRONMENT: Specified for 0°C to 50°C.
Specified to 80% R.H. at 35°C.
STORAGE ENVIRONMENT: –25°C to 65°C.
WEIGHT: 0.45kg (1 lb).
Channel 1
IN
OUT
Channel 40
IN
OUT
(Channels 2–39)
7705 40-Channel Control Module
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...