3-6
Routine Maintenance
5.
De-solder the blown CH21 or CH22 fuse as required, taking care not to damage the cir-
cuit board or spread solder flux around the board.
6.
Install a new 3A, 250V fast-blow fuse, Keithley part number FU-107-1.
CAUTION
Do not use a fuse with a higher current rating than specified or module
damage may occur.
7.
Solder the new fuse in place using organic (OA based) flux solder, again taking care not
to damage the circuit board or spread solder flux around the board.
8.
Carefully clean the repaired area of the circuit board with a foam tipped swab or brush
dipped in pure water, then blow dry the board with dry nitrogen gas. Allow the board to
dry for several hours in a 50˚C low-humidity environment before use.
9.
Re-install the circuit board into the plastic housing, then close the top cover.
Replacing Model 7702 amps fuses
1.
Turn off the power, and disconnect the power line and external connections from the
Model 7702.
2.
Open the Model 7702 top cover.
3.
Locate the amps fuses for CH41 and CH42 (see
Figure 3-4
).
4.
Remove the circuit board from the bottom plastic housing by removing the two bottom
screws.
5.
De-solder the blown CH41 or CH42 fuse as required, taking care not to damage the cir-
cuit board or spread solder flux around the board.
6.
Install a new 3A, 250V fast-blow fuse, Keithley part number FU-107-1.
CAUTION
Do not use a fuse with a higher current rating than specified or module
damage may occur.
7.
Solder the new fuse in place using organic (OA based) flux solder, again taking care not
to damage the circuit board or spread solder flux around the board.
8.
Carefully clean the repaired area of the circuit board with a foam tipped swab or brush
dipped in pure water, then blow dry the board with dry nitrogen gas. Allow the board to
dry for several hours in a 50˚C low-humidity environment before use.
9.
Re-install the circuit board into the plastic housing, then close the top cover.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...