![Keithley 2700 Service Manual Download Page 125](http://html1.mh-extra.com/html/keithley/2700/2700_service-manual_3944616125.webp)
Specifications
A-5
AC SPECIFICATIONS
1
Accuracy: ±(% of r % of range), 23°C ± 5°C
Calibration
3 Hz–
10 Hz–
20 kHz–
50 kHz–
100 kHz–
Function
Range
Resolution
Cycle
10 Hz
20 kHz
50 kHz
100 kHz
300 kHz
Voltage
2
100.0000 mV
0.1 µV
90 Days
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.6 + 0.08
4.0 + 0.5
1.000000
V
1.0 µV
10.00000
V
10 µV
1 Year
0.35 + 0.03
0.06 + 0.03
0.12 + 0.05
0.6 + 0.08
4.0 + 0.5
100.0000
V
100 µV
750.000
V
1.0 µV (
Temp. Coeff.)/°C
3
0.035 + .003
0.005 + .003
0.006 + .005
0.01 + .006
0.03 + .01
3 Hz–10 Hz
10 Hz–5 kHz
Current
2
1.000000
A
1.0 µA
90 Day/1 Yr.
0.30 + 0.04
0.10 + 0.04
3.00000
A
10 µA
0.35 + 0.06
0.15 + 0.06
(
Temp. Coeff.)/°C
3
0.035 + 0.006
0.015 + 0.006
(3 Hz–500 kHz) (333 ms–2 µs)
Frequency
4
100 mV
0.333 ppm
90 Day/ 1 Yr.
100 ppm + 0.333 ppm (SLOW, 1s gate)
and Period
to
3.33 ppm
100 ppm + 3.33 ppm (MED, 100ms gate)
750
V
33.3 ppm
100 ppm + 33.3 ppm (FAST, 10ms gate)
Additional Uncertainty ±(% of reading)
Low Frequency Uncertainty
MED
FAST
20 Hz - 30 Hz
0.3
—
30 Hz - 50 Hz
0
—
50 Hz - 100 Hz
0
1.0
100 Hz - 200 Hz
0
0.18
200 Hz - 300 Hz
0
0.10
>300 Hz
0
0
CREST FACTOR:
5
1 -2
2 - 3
3 - 4
4 - 5
Additional Uncertainty:
0.05
0.15
0.30
0.40
AC MEASUREMENT CHARACTERISTICS
AC Volts
MEASUREMENT METHOD: AC-coupled, True RMS.
INPUT IMPEDANCE: 1M
Ω
±2% // by <100pF.
INPUT PROTECTION: 1000Vp or 400VDC. 300Vrms with plug in modules.
AC Current
MEASUREMENT METHOD: AC-coupled, True RMS.
SHUNT RESISTANCE: 0.1
Ω
.
BURDEN VOLTAGE: 1A <0.3Vrms, 3A <1Vrms. Add 1Vrms when used with plug in modules.
INPUT PROTECTION: 3A, 250V fuse.
Frequency and Period
MEASUREMENT METHOD: Reciprocal Counting technique.
GATE TIME: SLOW 1s, MED 100ms, and FAST 10ms.
AC General
AC CMRR
6
: 70dB.
MAXIMUM CREST FACTOR: 5 at full-scale.
VOLT HERTZ PRODUCT: <= 8
×
10
7
.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...