2-14
Calibration
Short and open calibration
1.
Connect the Model 8610 low-thermal short to the instrument INPUT and SENSE termi-
nals as shown in
Figure 2-1
. Make sure the INPUTS button is not pressed in so that the
front inputs are active. Wait at least three minutes before proceeding to allow for ther-
mal equilibrium.
NOTE
Be sure to connect the low-thermal short properly to the HI, LO, and SENSE termi-
nals. Keep drafts away from low-thermal connections to avoid thermal drift, which
could affect calibration accuracy.
2.
Send the following command:
:CAL:PROT:DC:STEP1
3.
After the Model 2700 completes this step, remove the low-thermal short, and then send
this command:
:CAL:PROT:DC:STEP2
NOTE
Be sure to minimize movement near front Input terminals. Excessive movements can
cause capacitive coupling errors, which could affect calibration accuracy.
DC volts calibration
After the front panel short and open steps, do the following:
1.
Connect the calibrator to the Model 2700 as shown in
Figure 2-2
. Allow three minutes
for thermal equilibrium.
NOTE
Although 4-wire connections are shown, the sense leads are connected and discon-
nected at various points in this procedure by turning calibrator external sense on or
off as appropriate. If your calibrator does not have provisions for turning external
sense on and off, disconnect the sense leads when external sensing is to be turned off,
and connect the sense leads when external sensing is to be turned on.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...