Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-352
2606B-901-01 Rev. B / May 2018
trigger.timer[N].reset()
This function resets some of the trigger timer settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.timer[
N
].reset()
N
Trigger timer number (1 to 8)
Details
The
trigger.timer[
N
].reset()
function resets the following attributes to their factory defaults:
trigger.timer[
N
].count
trigger.timer[
N
].delay
trigger.timer[
N
].delaylist
trigger.timer[
N
].passthrough
trigger.timer[
N
].stimulus
It also clears
trigger.timer[
N
].overrun
.
Example
trigger.timer[1].reset()
Resets the attributes associated with timer 1 back
to factory default values.
Also see
(on page 7-348)
(on page 7-349)
(on page 7-350)
(on page 7-352)
trigger.timer[N].stimulus
This attribute specifies which event starts the timer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Not saved
0
Usage
eventID
= trigger.timer[
N
].stimulus
trigger.timer[
N
].stimulus =
eventID
eventID
The event that triggers the timer delay
N
Trigger timer number (1 to 8)