Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-190
2606B-901-01 Rev. B / May 2018
Example
smua.contact.speed = smua.CONTACT_SLOW
Configure contact check for higher accuracy
on SMU channel A.
Also see
(on page 2-50)
(on page 7-159)
(on page 7-187)
(on page 7-188)
(on page 7-210)
smuX.contact.threshold
This attribute stores the resistance threshold for the
smu
X
.contact.check()
function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall setup
Not saved
50 (50
Ω)
Usage
rValue
= smu
X
.contact.threshold
smu
X
.contact.threshold =
rValue
rValue
The resistance above which contact check should fail (measured in ohms)
X
Source-measure unit (SMU) channel (for example,
smua.contact.threshold
applies to SMU channel A)
Details
The threshold should be set to less than 1 kΩ.
Example
smua.contact.threshold = 5
Set the contact check threshold for SMU channel A to 5 Ω.
Also see
(on page 2-50)
(on page 7-159)
(on page 7-187)
(on page 7-210)