2606B System SourceMeter® Instrument Reference Manual
Appendix E: Status model
2606B-901-01 Rev. B / May 2018
E-11
Figure 149: Questionable status registers
Clearing registers
Commands to reset the status registers are listed in the table below.
In addition to these commands, you can reset the enable registers and the NTR to 0. To do this, send
the individual command to program the register with a 0 as its parameter value. The PTR registers
can be reset to their defaults by programming them with all bits on. Note that the event registers are
not programmable but can be cleared by reading them.
Commands to reset registers
Command
Description
To reset registers:
*CLS
Reset bits of the event and NTR registers to 0 and set all PTR register bits
on. Also clears the output queue.
status.reset()
Reset bits of the event and NTR registers to 0 and set all PTR register bits
on.