2606B System SourceMeter® Instrument Reference Manual
Section 9: Frequently asked questions
2606B-901-01 Rev. B / May 2018
9-5
Where can I get the LabVIEW driver?
The latest NI
TM
LabVIEW
TM
driver is available on
Why is the reading value 9.91e37?
This value indicates that there is a measurement overflow error. This error occurs when:
•
A measurement performed on a fixed range has a measured value greater than the specified
range
•
The measured value is larger than the maximum current or voltage range of the instrument
(exceeds the instrument rating)
If the instrument displays the overflow message on a particular range, select a higher range until an
on-range reading is displayed. To ensure the best accuracy and resolution, use the lowest range
possible that does not cause an overflow.