2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-141
localnode.password
This attribute stores the remote access password.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (W)
Yes
LAN reset
LAN restore defaults
Nonvolatile memory
""
Usage
localnode.password = "
password"
passWord
A string that contains the remote interface password
Details
This write-only attribute stores the password that is set for any remote interface. When password
usage is enabled (
localnode.passwordmode
), you must supply a password to change the
configuration or to control an instrument from a web page or other remote command interface.
The instrument continues to use the old password for all interactions until the command to change it
executes. When changing the password, give the instrument time to execute the command before
attempting to use the new password.
You cannot retrieve a lost password from any command interface.
You can reset the password by resetting the LAN from the virtual front panel or by using the
lan.reset()
command.
When using this command from a remote node,
localnode
should be replaced with the node
reference, for example,
node[5].password
.
Example
localnode.password = "N3wpa55w0rd"
Changes the remote interface password to
N3wpa55w0rd.
Also see
(on page 7-121)
(on page 7-141)
localnode.passwordmode
This attribute stores the remote access password enable mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
1 (localnode.PASSWORD_WEB)
Usage
mode
= localnode.passwordmode
localnode.passwordmode =
mode
mode
The remote password enable mode