Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-284
2606B-901-01 Rev. B / May 2018
Bit
Value
Description
B3
status.operation.instrument.smuX.trigger_overrun.MEAS
Set bit indicates that the
measurement event
detector of the SMU
was already in the
detected state when a
trigger was received.
Bit B3 decimal value: 8
B4
status.operation.instrument.smuX.trigger_overrun.ENDP
Set bit indicates that the
end pulse event
detector of the SMU
was already in the
detected state when a
trigger was received.
Bit B4 decimal value: 16
B5-B15
Not used
Not applicable.
As an example, to set bit B1 of the operation status SMU A trigger overrun enable register, set
status.operation.instrument.smua.trigger_overrun.enable
=
status.operation.instrument.smua.trigger_overrun.ARM
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B1 and B4, set
operationRegister
to 18 (which is the
sum of 2 + 16).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example
status.operation.instrument.smua.trigger_overrun.enable =
status.operation.instrument.smua.trigger_overrun.ARM
Uses a constant to sets
the ARM bit of the
operation status SMU A
trigger overrun enable
register.
Also see
(on page E-7)
status.operation.instrument.smuX.*
(on page 7-280)