2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-363
Example
print(tsplink.trigger[1].overrun)
If an event was ignored, displays
true
; if an event was not ignored,
displays
false
.
Also see
(on page 7-359)
(on page 7-360)
(on page 7-361)
(on page 7-363)
(on page 7-364)
(on page 7-365)
(on page 7-366)
tsplink.trigger[N].pulsewidth
This attribute sets the length of time that the trigger line is asserted for output triggers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
TSP-Link trigger N reset
Recall setup
Not saved
10e-6 (10 µs)
Usage
width
= tsplink.trigger[
N
].pulsewidth
tsplink.trigger[
N
].pulsewidth =
width
width
The pulse width (in seconds)
N
The trigger line (1 to 3)
Details
Setting the pulse width to 0 (seconds) asserts the trigger indefinitely.
Example
tsplink.trigger[3].pulsewidth = 20e-6
Sets pulse width for trigger line 3 to 20
μs.
Also see
(on page 7-363)
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[
N
].release()
N
The trigger line (1 to 3)