2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-255
As an example, to set bit B1 of the measurement event buffer available summary enable register, set
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
.
In addition to the above constants,
measurementRegister
can be set to the decimal equivalent of
the bit to set. To set more than one bit of the register, set
measurementRegister
to the sum of
their decimal weights. For example, to set bits B1 and B2, set
measurementRegister
to 6 (which is
the sum of 2 + 4).
Example
status.measurement.buffer_available.enable =
status.measurement.buffer_available.SMUA
Sets the SMUA bit of the measurement event buffer available summary enable register.
Also see
(on page E-21)
status.measurement.current_limit.*
This attribute contains the measurement event current limit summary registers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
6 (All bits set)
Usage
measurementRegister
= status.measurement.current_limit.condition
measurementRegister
= status.measurement.current_limit.enable
measurementRegister
= status.measurement.current_limit.event
measurementRegister
= status.measurement.current_limit.ntr
measurementRegister
= status.measurement.current_limit.ptr
status.measurement.current_limit.enable =
measurementRegister
status.measurement.current_limit.ntr =
measurementRegister
status.measurement.current_limit.ptr =
measurementRegister
measurementRegister
The status of the measurement event current limit summary register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various
bit settings
Details
These attributes are used to read or write to the measurement event current limit summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.
For example, assume value 6 is retu+rned for the enable register. The binary equivalent is
0000 0000 0000 0110. This value indicates that bit B1 (SMUA) and bit B2 (SMUB) are set.