2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-237
smuX.trigger.measure.stimulus
This attribute selects the event that causes the measure event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0
Usage
eventID
= smu
X
.trigger.measure.stimulus
smu
X
.trigger.measure.stimulus =
eventID
eventID
Event that triggers the measurement detector
X
Source-measure unit (SMU) channel (for example,
smua.trigger.measure.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event. When set, the
SMU waits for the event at the measurement event detector portion of the trigger model.
Set this attribute to zero to bypass waiting for an event (the SMU continues uninterrupted through the
remote trigger model). Set
eventID
to one of the existing trigger event IDs shown in the following
table.
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU) transitions
from the idle state to the arm layer of the trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to the
trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement action
smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smu
X
.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[
N
].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object
N
display.trigger.EVENT_ID
Occurs when the TRIG key on the virtual front panel is
pressed
trigger.EVENT_ID
Occurs when a
*TRG
command is received on the
remote interface
USB only: Occurs when a USBTMC TRIGGER message
is received
VXI-11 only: Occurs with the VXI-11 command
device_trigger
; reference the VXI-11 standard for
additional details on the device trigger operation
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected