2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-133
lan.trigger[N].ipaddress
This attribute specifies the address (in dotted-decimal format) of UDP or TCP listeners.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
LAN trigger N reset
Recall setup
Not saved
"0.0.0.0"
Usage
ipAddress
= lan.trigger[
N
].ipaddress
lan.trigger[
N
].ipaddress =
ipAddress
ipAddress
The LAN address for this attribute as a string in dotted decimal notation
N
The LAN event number (1 to 8)
Details
Sets the IP address for outgoing trigger events.
Set to
"0.0.0.0"
for multicast.
After changing this setting, the
lan.trigger[
N
].connect()
command must be called before
outgoing messages can be sent.
Example
lan.trigger[3].protocol = lan.TCP
lan.trigger[3].ipaddress = "192.168.1.100"
lan.trigger[3].connect()
Set the protocol for LAN trigger 3 to be
lan.TCP
when sending LAN triggers.
Use IP address
"192.168.1.100"
to
connect the LAN trigger.
Also see
lan.trigger[N].mode
This attribute sets the trigger operation and detection mode of the specified LAN event.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
LAN trigger N reset
Recall setup
Not saved
0 (lan.TRIG_EITHER)
Usage
mode
= lan.trigger[
N
].mode
lan.trigger[
N
].mode =
mode
mode
A number representing the trigger mode (0 to 7); see the
Details
section for
more information
N
A number representing the LAN event number (1 to 8)
Details
This command controls how the trigger event detector and the output trigger generator operate on the
given trigger. These settings are intended to provide behavior similar to the digital I/O triggers.