![background image](http://html1.mh-extra.com/html/keithley/2606b-system-sourcemeter/2606b-system-sourcemeter_reference-manual_4051272690.webp)
Appendix B: Calibration
2606B System SourceMeter® Instrument Reference Manual
B-6
2606B-901-01 Rev. B / May 2018
Current source accuracy
To verify that the 2606B output current accuracy is within specified limits:
1. With the power off, connect the digital multimeter to the 2606B terminals as shown in the figure
below.
Figure 131: Calibration connections for 100 nA to 1 A current ranges
Figure 132: Calibration connections for 3 A current range