Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-228
2606B-901-01 Rev. B / May 2018
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU)
transitions from the idle state to the arm layer of the
trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to
the trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement
action
smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smu
X
.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a digital I/O
line
tsplink.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[
N
].EVENT_ID
Occurs when the appropriate LXI trigger packet is
received on LAN trigger object
N
display.trigger.EVENT_ID
Occurs when the TRIG key on the virtual front panel
is pressed
trigger.EVENT_ID
Occurs when a
*TRG
command is received on the
remote interface
USB only: Occurs when a USBTMC TRIGGER
message is received
VXI-11 only: Occurs with the VXI-11 command
device_trigger
; reference the VXI-11 standard
for additional details on the device trigger operation
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected
trigger.timer[
N
].EVENT_ID
Occurs when a delay expires
trigger.generator[
N
].EVENT_ID
Occurs when the
trigger.generator[
N
].assert()
function is
executed
* Use the name of the trigger event ID to set the stimulus value rather than the numeric value. Using the name
makes the code compatible for future upgrades (for example, if the numeric values must change when
enhancements are added to the instrument).
Example
smua.trigger.arm.stimulus = trigger.timer[1].EVENT_ID
An event on trigger timer 1 causes the arm event detector to enter the detected state.
Also see
(on page 3-30)