Appendix B: Calibration
2606B System SourceMeter® Instrument Reference Manual
B-20
2606B-901-01 Rev. B / May 2018
Typical values for the 1 V range:
smua.source.calibratev(-1, -1e-30, -1e-4, -0.9, -0.896)
22. If this step is not on the CALA sense mode, send the measure calibration command using the
multimeter and 2606B readings and range setting for the parameters:
smua.measure.calibratev(
-range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg
)
Where:
-range
= The negative of the present calibration range
Z_rdg
= The -zero 2606B measurement
DMM_Z_rdg
= The -zero DMM measurement
FS_rdg
= The -FS 2606B measurement
DMM_FS_rdg
= The -FS DMM measurement
Typical 1 V range values:
smua.measure.calibratev(-1, -1e-4, -1e-6, -0.89, -0.896)
23. Select automatic polarity mode:
smua.cal.polarity = smua.CAL_AUTO
Current calibration adjustment
1. Connect the 2606B SMU to the digital multimeter (see the following figure), and then select the
multimeter DC current function.
Figure 135: Calibration connections for 100 nA to 1 A current ranges
2. Send this command to initialize current calibration:
smua.source.func = smua.OUTPUT_DCAMPS