Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-350
2606B-901-01 Rev. B / May 2018
Also see
(on page 7-352)
trigger.timer[N].EVENT_ID
This constant specifies the trigger timer event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= trigger.timer[
N
].EVENT_ID
eventID
The trigger event number
N
Trigger timer number (1 to 8)
Details
This constant is an identification number that identifies events generated by this timer.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
events from this timer.
Example
trigger.timer[1].stimulus = tsplink.trigger[2].EVENT_ID
Sets the trigger stimulus of
trigger timer 1 to the TSP-
Link trigger 2 event.
Also see
None
trigger.timer[N].overrun
This attribute indicates if an event was ignored because of the event detector state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Instrument reset
Recall setup
Trigger timer N clear
Trigger timer N reset
Not applicable
false
Usage
overrun
= trigger.timer[
N
].overrun
overrun
Trigger overrun state (
true
or
false
)
N
Trigger timer number (1 to 8)
Details
This command indicates if an event was ignored because the event detector was already in the
detected state when the event occurred.