2606B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2606B-901-01 Rev. B / May 2018
7-351
This is an indication of the state of the event detector built into the timer itself. It does not indicate if
an overrun occurred in any other part of the trigger model or in any other construct that is monitoring
the delay completion event. It also is not an indication of a delay overrun.
Delay overrun indications are provided in the status model.
Example
print(trigger.timer[1].overrun)
If an event was ignored, the output
is
true
.
If the event was not ignored, the
output is
false
.
Also see
(on page 7-352)
trigger.timer[N].passthrough
This attribute enables or disables the timer trigger pass-through mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Not saved
false (disabled)
Usage
passthrough
= trigger.timer[
N
].passthrough
trigger.timer[
N
].passthrough =
passthrough
passthrough
The state of pass-through mode; set to one of the following values:
true
: Enabled
false
: Disabled
N
Trigger timer number (1 to 8)
Details
When pass-through mode is enabled, triggers are passed through immediately and initiate the delay.
When disabled, a trigger only initiates a delay.
Example
trigger.timer[1].passthrough = true
Enables pass-through mode on trigger timer 1.
Also see
(on page 7-352)