9 Technical data
INSTALLATION REQUIREMENTS & ENVIRONMENT CONTROL:
In order to minimize interference risks, the following requirements shall apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must meet the requirements given in
on page 87. Use of incorrect cables may result in the device causing radio frequency
interference.
Electrostatic discharges environment & recommendations
In order to reduce electrostatic discharge interference, a charge dissipative floor should be installed to
prevent charge accumulation.
• The dissipative floor material must be connected to the system reference ground, if applicable.
• Relative humidity must be maintained above 30 percent.
Stacked components & equipment
WARNING!
The PCX-1 should not be used adjacent to or stacked with other equipment; if
adjacent or stacked use is necessary, the PCX-1 should be observed to verify normal operation in
the configuration in which it will be used.
Interference may occur in the vicinity of equipment marked with the following symbol:
9.7
X-ray tube assemblies
ORTHOPANTOMOGRAPH
™
OP 3D
105