7 Maintenance
7.2.3.4
PAN pixel calibration (2D sensor)
This program calibrates the 2D sensor for Panoramic imaging.
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
1.
Select
PAN pixel calibration (2D sensor)
from the Calibrations menu.
2.
Protect yourself from radiation.
3.
Press and hold the exposure button down to take the calibration image.
4.
When the exposure warning stops and the program end tone is played, the program is complete.
5.
The calibration image appears on the GUI.
6.
Acknowledge the calibration result by pressing the
OK
button.
7.2.3.5
CEPH pixel calibration
This program calibrates the sensor for Cephalometric imaging.
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
1.
Turn the CEPH head support to LAT position.
2.
Move the nasion support to its outmost position an turn it down.
3.
Select
CEPH pixel calibration
from the Calibrations menu.
4.
Protect yourself from radiation.
5.
Press and hold the exposure button down to take the calibration image.
6.
When the exposure warning stops and the program end tone is played, the program is complete.
76
ORTHOPANTOMOGRAPH
™
OP 3D