9 Technical data
Device data
Model:
PCX–1
Protection against electric shock
Class I
Degree of protection
Type B applied with no conductive connection to the patient
Protection against the ingress of liquids IP20
Cleaning agents and protection against
cross contamination
• Distilled water
• Ethanol 96%
• Isopropyl alcohol
• Soapy water
• Cavicide
™
disinfectant
• Metasys
™
disinfectant
• Disposable plastic covers for bite block, chin rest, lip
support and head support
• Disposable plastic covers for ear rods and nasion support
(CEPH only)
Use environment
In environments where no flammable anesthetics nor
flammable cleaning agents are present
Mode of operation
Continuous operation/intermittent loading
Safety
IEC60601–1
Source of power
Mains connection
EMC Classification
Class B
Tubehead assembly
Tubehead assembly
THA/HVGEN PCX
Tube type
D–054S or equivalent
Stationary anode
Tube voltage
60 – 95 kV
Max. tube current
16 mA
Target angle
5°
Focal spot
0.5 (IEC 60336/2005)
Total filtration
min. 3.2 mm Al @ 95 kV
Additional filtration of cover materials
(CEPH only)
max. 1.6 mm Al @ 95 kV
Leakage technique factors
PAN: 4613 mAs/h @ 90 kV/12.5 mA
3D: 2880 mAs/h @ 95 kV/4 mA
CEPH: 4736 mAs/h @ 95 kV/11 mA
NOTICE!
Leakage technique factors may be impacted by ambient temperature or used
workstation.
88
ORTHOPANTOMOGRAPH
™
OP 3D