7 Maintenance
15.
Detach the copper filter from the CEPH tubehead if it was used.
7.3.3
3D QC
NOTICE!
This program is available only for devices with 3D modality.
1.
GUI:
Go to device settings.
2. GUI:
Select
3D QC
program from the
Quality control
menu.
3.
Attach the 3D QC phantom holder and the 3D QC phantom to the device.
4.
Protect yourself from radiation.
5.
Press and hold the exposure button down to take the QC image.
6.
The QC image preview appears on the GUI, showing the result of the check.
7.
Acknowledge the preview by pressing the
OK
button on the GUI.
82
ORTHOPANTOMOGRAPH
™
OP 3D