3 Overview
DENTAL CAST IMAGING TOOLS (OPTIONAL):
Dental cast holder
3.5
Emergency stop switch
An emergency stop switch is located on the left side of the carriage.
Pressing the emergency stop switch immediately terminates the imaging and all device movements.
NOTICE!
An interrupted imaging process cannot be resumed. A new image needs to be taken.
To release the emergency stop switch, rotate it clockwise.
NOTICE!
Make sure that the emergency stop switch is not pressed down when you start the
imaging process.
The Graphical User Interface (GUI) has indication if the emergency stop is active and the device status
indication lights are blue.
NOTICE!
Activating the emergency stop causes the device to restart.