![KaVo ORTHOPANTOMOGRAPH OP 3D User Manual Download Page 61](http://html2.mh-extra.com/html/kavo/orthopantomograph-op-3d/orthopantomograph-op-3d_user-manual_3754897061.webp)
6 Using the device
6.4.3
Patient positioning for Carpus imaging
The Carpus imaging program requires a special carpus holder (optional) to be used.
CAUTION!
Before taking a Carpus image, make sure the imaging method is approved by local
authorities.
1.
Press
HOME
button on the patient positioning panel. The device moves to CEPH HOME (Patient-In)
position.
NOTICE!
Ensure that the device has been set up correctly for the required examination.
2.
The device goes to ready state and the status indicator light turns green.
NOTICE!
If the status indicator light does not turn green, look for more information on the
GUI.
3.
Turn the cephalometric unit's head support to PA position.
4.
Move the nasion support to its outmost position and turn it aside, to the direction shown in the
image. You may also turn the ear rods aside for easier positioning.
ORTHOPANTOMOGRAPH
™
OP 3D
61