6 Using the device
5.
Attach the carpus holder by sliding it to the base of the nasion support.
6.
Adjust the device's height if needed.
7.
Ask the patient to remove any jewellery and all other things that may cause artifacts to the image.
8.
Protect the patient from radiation according to the local regulations, for example using a lead
apron.
9.
Ask the patient to place their palm against the carpus holder with their fingers spread apart and
wrist straightened. The palm of the hand should face the sensor.
NOTICE!
Only the center area of the carpus holder is exposed. Ensure that the patient's hand
is positioned so that the imaged parts are on the center area of the carpus holder (white
area in the image above).
NOTICE!
You may place the patient's hand diagonally on the carpus holder if it doesn't fit
properly otherwise.
NOTICE!
To ensure optimal image quality, pay attention to the correct patient positioning.