7 Maintenance
3.
Attach the PAN QC phantom holder and 2D QC test phantom to the device.
4.
Attach the copper filter in front of the radiation window on the PAN/3D tubehead. The filter
attaches in place with magnets.
5.
Protect yourself from radiation.
6.
Press and hold the exposure button down to take the QC image.
7.
The QC image preview appears on the GUI.
8.
Acknowledge the preview by pressing the
OK
button on the GUI.
78
ORTHOPANTOMOGRAPH
™
OP 3D