9 Technical data
Electromagnetic immunity IEC 60601-1-2 Ed4
PCX-1 is suitable for use in the specified electromagnetic environment. The purchaser or user of PCX-1 should assure that it is
used in an electromagnetic environment as described below:
Immunity Test
IEC60601-1-2 Test
Level
Compliance Level
Electromagnetic Environment
Electrostatic discharge
(ESD) IEC61000-4-2
± 8 kV for contact
discharge
±2, 4, 8, 15 kV for air
discharge
± 8 kV for contact
discharge
±2, 4, 8,15 kV for air
discharge
Floors are wood, concrete, or ceramic tile, or
floors are covered with synthetic material and
the relative humidity is at least 30 percent.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power supply
lines (100 kHz)
±1 kV for input/output
lines (100kHz)
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality is that of a typical
commercial and/or hospital environment
Surge
IEC61000-4-5
± 0.5, 1 kV differential
mode
± 0.5, 1, 2 kV common
mode
± 0.5, 1 kV differential
mode
± 0.5, 1, 2 kV common
mode
Mains power quality is that of a typical
commercial and/or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC61000-4-11
-0 %
U
mains; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
-0 %
U
mains; 1 cycle, At
0°
-70 %
U
mains; 25/30
cycles At 0°
-0 %
U
mains; 250/300
cycle At 0°
-0 %
U
mains; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
-0 %
U
mains; 1 cycle, At
0°
-70 %
U
mains; 25/30
cycles At 0°
-0 %
U
mains; 250/300
cycle At 0°
Mains power quality is that of a typical
commercial and/or hospital environment.
If the user of
PCX-1
requires continued
operation during power mains interruptions, it is
recommended that
PCX-1
be powered from an
uninterruptible power supply.
Power frequency
(50/60Hz) magnetic
field IEC 61000-4-8
30A/m
30 A/m
Power frequency magnetic fields are at levels
characteristic of a typical location in a typical
commercial and/ or hospital environment.
The device shall not be used closer than 15cm to
sources of 50/60Hz magnetic field.
NOTICE!
U
T
is the a.c. mains voltage prior to application of the test level.
102
ORTHOPANTOMOGRAPH
™
OP 3D