9 Technical data
Electrical connections
Nominal mains voltage
100 - 240 VAC
Tolerance: ± 10%
Input power frequency
50 / 60 Hz
Nominal current
10 A @ 220–240 VAC, 15 A @ 100 – 120 VAC
Main fuses (F1 & F2)
220–240 VAC:
Littelfuse 215 (Time–Lag) 10 A
Cooper Bussman (Time Delay) S505H–10–R
100–120 VAC:
Littelfuse 326 (SlowBlow) 15 A
Cooper Bussman (Time Delay) MDA–15
External warning light fuse (F3)
Cooper Bussmann (Time Delay) S506–2–R 2 A
Power consumption
2.3 kVA @ 220–240 VAC, 1.65 kVA @ 100–120 VAC
Maximum impedance of mains
0.2 Ω
Recommended circuit breaker
max. 30 mA
Mains cord
EU 230V: 16 A / 250 VAC; 50/60 Hz H05VV–F3G1.5
US 115V: 15 A / 125 VAC 50/60 Hz SJT 3 x AWG 14
US 230V: 15 A / 250 VAC 50/60 Hz SJT 3 x AWG 14
Exposure controller cable
Remote exposure switch: DINFLEX–YY 4 x AWG 26, max. 15
m
Data communication cable
CAT6 UTP Ethernet cable
Positioning laser lights
Panoramic laser lights (3)
Cephalostat laser lights (1)
3D laser lights (3)
IEC 60825–1/2001 (CLASS 1 LASER PRODUCT)
X–ray generator
Generator
High frequency DC generator
Anode voltage
60 – 95 kV (± 5 kV)
Anode current
2 – 16 mA (± 20%)
Exposure time accuracy
± 5%
DAP display accuracy
± 25%
Spine compensation
kV / mA compensated
Accuracy of radiation output
> 95%
ORTHOPANTOMOGRAPH
™
OP 3D
89