9 Installation
Configurations for service personnel:
Calibration
intervals
Change the required calibration and QC
interval for the User and for Service
(full calibration) according to the local
regulations. The device will remind the user
when re-calibration and new PAN QC and 3D
QC retakes are required according to this
setting.
The default values are:
• User interval - Biannually
• Service interval - Annually
• PAN QC interval - Quarterly
• 3D QC interval - Quarterly
License
If the user has purchased a licensable
upgrade for optional imaging programs, for
example 9x14 3D FOV, they can be activated
through this field.
Type the purchased license key to the field
and press Save to activate the option.
Select country
Select the installation country of the device.
The selection configures the device according
to country specific regulations, for example
change the maximum selectable mA for
imaging.
NOTICE!
For PAN QC to be used with
1.8mm Cu filter, select Germany.
Door Switch
If an optional Door Switch has been installed
to the device according to chapter
Back
panel connectors
, it has to be activated here.
Demo mode
Used for setting the device into
demonstration mode where the radiation
production of the device is disabled while
maintaining all other functionalities.
NOTICE!
Demo mode is intended for
exhibition use only.
DAP correction
Change the DAP correction factor which is
used to calculate the shown DAP value for
the currently selected imaging program. This
setting is preset at the factory and should
not be altered in normal conditions.
NOTICE!
The DAP correction factor
has no impact on the device radiation
output, it is used for DAP display
calibration purposes only.
Configurations
PAN QC spinal
compensation
Enable/Disable spinal compensation for PAN
QC program.
ORTHOPANTOMOGRAPH
™
OP 3D