7 Maintenance
7.3.3 Calibration procedure
Perform the calibrations in the exact order instructed in here.
3D Pixel Calibration
This program calibrates the sensor for 3D imaging.
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
1.
Select
3D Pixel calibration
from the Calibrations menu.
2.
Protect yourself from radiation.
3.
Press and hold down the exposure button to take the calibration image.
4.
When the exposure warning stops and the program end tone is played, the program
is complete.
5.
The calibration image appears on the GUI.
6.
Acknowledge the calibration result by pressing the
OK
button.
PAN Pixel Calibration
This program calibrates the sensor for Panoramic imaging.
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
1.
Select
PAN Pixel Calibration
from the Calibrations menu.
2.
Protect yourself from radiation.
3.
Press and hold down the exposure button to take the calibration image.
50
ORTHOPANTOMOGRAPH
™
OP 3D