9 Installation
9.9.2 Mains power configuration
Detach the indicated screw on the back panel cover and lift it off to access the device
back panel.
Mains connection type configuration
IEC 60601-2-63 standard requires extraoral X-ray devices to be installed permanently.
The device can, however, be installed with plug connection if the local regulations allow
this. In these cases, order a mains cord with plug from the manufacturer.
The i4100 Back Panel Board contains a switch (SW1), which needs to be configured if the
type of installation is changed from permanent to plug connection.
NOTICE!
If device has been delivered from the factory with plug connection, it
has already been configured accordingly.
To access the switch, you must detach the Back panel plate by removing the two
indicated screws.
Mains power configurations:
Permanent connection
- Switch (SW1) shall be in
F1
position.
Plug connection
- Switch (SW1) shall be in
F2
position and both fuses (F1+F2) must
be installed.
See chapter
Changing the fuses
for information on how to install the fuses.
Mains voltage configuration
The device can be used in 220-240 VAC or 100-120 VAC mains voltage. The installed
main fuses (F1+F2) need to be configured according to the mains voltage:
• 10A suited for 220-240 VAC mains voltage
• 15A suited for 100-120 VAC mains voltage