9 Installation
9.8.7 Collimator Calibration
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
NOTICE!
The X-ray beam size is verified during this calibration.
1.
Select
Collimator Calibration
from the Calibrations menu.
2.
Protect yourself from radiation.
3.
Press and hold down the exposure button to take the calibration image.
4.
When the exposure warning stops and the program end tone is played, the program
is complete.
5.
The calibration image appears on the GUI.
NOTICE!
If the collimator calibration fails and the GUI informs you to adjust
the height or tilt of the collimator, adjust the collimator according to chapter
Collimator adjustment
.
6.
Acknowledge the calibration result by pressing the OK button.
NOTICE!
Repeat the 3D Pixel Calibration and Device Geometry Calibration if any
mechanical adjustment was required.
9.8.8 PAN Pixel Calibration
This program calibrates the sensor for Panoramic imaging.
No calibration tools are required when performing this calibration.
NOTICE!
This calibration produces X-rays. Protect yourself from radiation.
1.
Select
PAN Pixel Calibration
from the Calibrations menu.
2.
Protect yourself from radiation.
ORTHOPANTOMOGRAPH
™
OP 3D
77