9 Installation
4.
Press and hold down the exposure button to take the calibration image.
5.
The calibration results appear on the GUI.
NOTICE!
This will take several minutes.
6.
Acknowledge the calibration by pressing the
OK
button.
7.
Repeat the calibration for the remaining FOV sizes.
9.8.10 PAN QC (optional)
NOTICE!
The PAN QC is an optional, yet recommendable, procedure and is
mandatory to be performed where local regulations require it.
NOTICE!
The device's country selection needs to be configured per local
requirements before performing the QC programs. See chapter
Device
configuration
.
1. GUI:
Go to Settings menu.
2. GUI:
Select
Quality Control
menu.
3. GUI:
Select
PAN QC
program from the list.
4.
Attach the Panoramic QC phantom holder and 2D QC phantom to the device.