10 Technical data
Electrical connections
Nominal mains voltage
100–240 Vac
Input power frequency
50 / 60 Hz
Nominal current
10A @ 220–240 VAC, 15A @ 100 – 120
VAC
Main fuses (F1 & F2)
220–240Vac:
Littelfuse 215 (Time–Lag) 10A
Cooper Bussman (Time Delay) S505H–10–
R
100–120Vac:
Littelfuse 326 (SlowBlow) 15A
Cooper Bussman (Time Delay) MDA–15
External warning light fuse (F3)
Cooper Bussmann (Time Delay) S506–2–R
Power consumption
2,3 kVA @ 220–240 VAC, 1,65 kVA @
100–120 VAC
Maximum impedance of mains
0,2 Ω
Recommended circuit breaker
max. 30 mA
Mains cord
EU: 16 A / 250 VAC; 50/60 Hz H05VV–
F3G1.5
US: 15 A / 125 VAC 50/60 Hz SJT 3 x AWG
14
Exposure controller cable
Exposure switch: Li12Y11Y 2 x AWG 26,
3m
Remote exposure switch: DINFLEX–YY 4 x
AWG 26, max. 15m
Data communication cable
CAT6 UTP Ethernet cable
Positioning laser lights
Panoramic laser lights (3)
3D laser lights (3)
Laser light
(CLASS 1 LASER PRODUCT)
IEC 60825–1/2001
The device contains warning symbol of the
laser lights.
100
ORTHOPANTOMOGRAPH
™
OP 3D