10 Technical data
Device data
Model:
PCX–1
Protection against electric shock
Class I
Degree of protection
Type B applied with no conductive
connection to the patient
Protection against the ingress of liquids
IP20
Cleaning methods
• Distilled water
• Ethanol 96%
• Isopropyl alcohol
• Soapy water
• Cavicide
™
disinfectant
• Metasys
™
disinfectant
• Disposable plastic covers for bite block,
chin rest, lip support and head support
Use environment
In environments where no flammable
anesthetics nor flammable cleaning agents
are present
Mode of operation
Continuous operation/intermittent loading
Safety
IEC60601–1
Source of power
Mains connection
EMC Classification
Class B
Tube head assembly
Tube head assembly
THA/HVGEN PCX
Tube type
Toshiba D–054S or equivalent
Stationary anode
Tube voltage
60 – 95 kV
Max. tube current
16 mA
Target angle
5°
Focal spot
0,5 (IEC 60336/2005)
Total filtration
min. 3,2mm Al @ 95kV
Leakage technical factors
2717 mAs/h @ 95kV/4mA
ORTHOPANTOMOGRAPH
™
OP 3D
99