9 Installation
8.
Re-install the wall bracket rail cover back in place.
9.
Re-install the cable clamp to its place.
10.
Continue to install the column normally according to chapter
Installing the device
.
9.9.5 Exhibition stand installation
If the device cannot be bolted to the floor, it may be installed on a special exhibition
stand (ordered separately).
NOTICE!
The device has to be fixed to the wall even when the exhibition stand is
used.
1.
Transport the exhibition stand to the location, where the device is to be installed.
2.
Attach the device column on the exhibition stand. See the installation instructions
supplied with the exhibition stand for more details.
3.
Complete the installation according to the normal installation instructions in chapter
Installing the device
.
9.9.6 Collimator adjustment
If the collimator calibration has failed due to the collimator being too high/low or tilted,
adjustment is needed.
NOTICE!
This adjustment is required only in special cases. After the adjustment,
all calibrations related to radiation must be performed again.
1.
Remove the tubehead rear cover by detaching the 3 indicated screws and pull the
cover outwards. The screw on the middle of the cover has a screw cover cap over it.
88
ORTHOPANTOMOGRAPH
™
OP 3D