9 Installation
20.
Pull the protective plastic aside from the carriage to reveal the base plate and back
panel connectors.
21.
Check that the device's mains power configuration is applicable to the installation
location.
NOTICE!
The device is delivered with 10A fuses installed. If installing to other
than 230V mains voltage the fuses need to be changed. See chapters
Mains
power configuration
and
Changing the fuses
for more information.
NOTICE!
If mains connection type needs to be changed, see chapter
Mains
power configuration
for more information.
22.
Move the carriage against the column.
NOTICE!
If the device is installed in an angle (e.g. corner installation) the
plywood supports may come in the way of the installation and need to be
removed at this point. See more details in chapter
Device dimensions
.
23.
Loosen the three carriage locking bolts on top of the column and ensure that they
are facing outwards.
24.
Push the carriage installation rail into the slot on the column. The rail locks itself into
the slot.
ORTHOPANTOMOGRAPH
™
OP 3D
61