9 Installation
12.
Save the passed QC image as a reference image for the user and let the user know
the reference image location.
9.8.11 3D QC
NOTICE!
The device's country selection needs to be configured per local
requirements before performing the QC programs. See chapter
Device
configuration
.
1. GUI:
Go to Settings menu.
2. GUI:
Select
Quality Control
menu.
3. GUI:
Select
3D QC
program from the list.
4.
Attach the 3D QC phantom holder and the 3D QC phantom to the device.
5.
Protect yourself from radiation.
6.
Press and hold down the exposure switch to take the QC image.
7.
The QC image preview appears on the GUI.
8.
Acknowledge the result by pressing the
OK
button on the GUI.
ORTHOPANTOMOGRAPH
™
OP 3D
81