10 Technical data
Scout programs & technical factors
NOTICE!
Tube voltage is fixed to 95 kV
FOV size
Resolution Pixel size mA range
Exposure
time
DAP range
5 x 5
Scout
200 μm
4.0 - 12.5
mA
0.04 - 0.03 s 4 - 9 mGycm
2
6 x 9
Scout
200 μm
4.0 - 12.5
mA
0.08 - 0.07 s 7 - 20 mGycm
2
9 x 11
Scout
200 μm
4.0 - 12.5
mA
0.08 - 0.07 s 7 -20 mGycm
2
9 x 14
(optional)
Scout
200 μm
4.0 - 12.5
mA
0.08 - 0.07 s 7 - 20 mGycm
2
3D imaging programs & technical factors
NOTICE!
Tube voltage is fixed to 95 kV
FOV size
Resolution Voxel size mA range
Exposure
time
DAP range
LDT
280 μm
2.0 - 4.0 mA 1.49 - 1.40 s 33 - 63 mGycm
2
Standard
200 μm
4.0 - 12.5
mA
3.06 - 2.87 s 137 - 402
mGycm
2
High
125 μm
2.0 - 4.0 mA 9.99 s
224 - 448
mGycm
2
5 x 5
ENDO
80 μm
2.0 - 4.0 mA 20.0 s
448 - 895
mGycm
2
LDT
320 μm
2.0 - 4.0 mA 4.48 - 4.19 s 121 - 227
mGycm
2
Standard
300 μm
4.0 - 12.5
mA
4.19 - 3.61 s 227 - 610
mGycm
2
6 x 9
High
200 μm
2.0 - 4.0 mA 20.0 s
541 - 1081
mGycm
2
LDT
320 μm
2.0 - 4.0 mA 4.48 - 4.19 s 211 - 394
mGycm
2
Standard
300 μm
4.0 - 12.5
mA
4.19 - 3.61 s 394 - 1060
mGycm
2
9 x 11
High
200 μm
2.0 - 4.0 mA 20.0 s
940 - 1879
mGycm
2
LDT
400 μm
2.0 - 4.0 mA 4.48 - 4.19 s 224 - 417
mGycm
2
Standard
350 μm
4.0 - 12.5
mA
4.19 - 3.61 s 419 - 1128
mGycm
2
9 x 14
(optional)
High
250 μm
2.0 - 4.0 mA 20.0 s
1000 - 2000
mGycm
2
ORTHOPANTOMOGRAPH
™
OP 3D
103